HOW DOES IT WORK?

ichroma™ II is an automatic or semiautomatic in-vitro diagnostic device that measures the concentration of analytes, contained in blood, urine, or other samples, in quantitative or semi-quantitative ways. Compared to the first-generation device, ichroma™, ichroma™ II has an improved user interface and display and an advanced optical system, which provide the user with disease test options in wide range.
In addition, ichroma™ II was designed to improve data input efficiency and wireless communication connectivity through separately purchasable accessories and program update can be proceeded easily using SD card. In particular, the use of batteries improves portability, which is one of the biggest advantages of point-of-care testing (POCT) equipment, and ensures reliable testing in unstable power environments.

STRUCTURE

MOVIE

ADVANTAGE

ADVANTAGE
Interface Expansion of user interface through adoption of Android OS
Display 7 "touch panel provides operational stability and quick response
Portability Battery for unexpected power failure or poor power environment for a certain period
Data storage and update Available to store and search large amount of test results
Data back-up and OS update using SD card
Optical technology Improved fluorescence optical module for high sensitive diagnosis
Measurement modes Single, multi-mode, QC test and review (save results) mode
Stability Reduced risk of test error based on the equation saving function
Connectivity Built-in printer, improved wired / wireless connectivity, LIS / HIS communication
USB hub chip supporting 4 USB ports
Convenience Keyboard adaptable for non-touchpad user
Improved convenience to input sample ID and tester ID

SPECIFICATIONS

ADVANTAGE
Dimensions 276 mm (L) x 220 mm (W) x 91 mm (H)
Weight 1.3 kg
Power supply DC 12V/5A, AC/DC Adaptor, (AA battery) DC 1.5V X 4ea. Input 100-240V ~ 50/60Hz, 1.8A. Output DC 12V/5A
Data output 7” touch color LCD
Motor Stepper Linear Actuator (5V, 0.34A)
Communication ports USB 4 ports, LAN port, USB OTG port
Operation temperature/ humidity range 15 ~ 35 °C
Storage temperature/humidity range Temperature -20 ~ 50°C, Humidity Max. 70%
Data output LCD and built-in printer
Test result storage Patient test results (1,000) / System check results (500)
Standard material (control) test results (1,000)
Save of ID chip information 500 ID chips (Max. 5 LOTs per each item)

FAQ

  • QCan I use whole blood in addition to plasma or serum samples?

    Yes. Whole blood is usable for specific items.

  • QCan I touch the touchscreen with latex gloves on?

    Yes. LCD touchscreen of ichroma™ II is pressure sensitive so, you may use ichroma™ II without taking off gloves

  • QWhat is the significant difference from ichroma™ reader?

    Compared to ichroma™ reader, user convenience (LCD touch screen, built-in printer, etc.) has been improved and Optical technology enabling high sensitivity test including infection marker is applied.